Cleaning attention of Laparoscopic Surgical Instruments

Mar 13, 2018
[[TitleContent]]

Cleaning attention of Laparoscopic Surgical Instruments

Thorough cleaning is a necessary condition to ensure disinfection and sterilization, and the cleaning of endoscopy and equipment is carried out in a special endoscope cleaning center. The Laparoscopic Surgical Instruments can remove the part should be fully open, the outer sheath separating forceps, scissors removed, rinsed with water cannons, with a soft brush repeatedly separated joints and occlusal surface to surface dirt; remove the puncture and switch on the rubber cap, a rubber flap, washed clean; puncture completely remove the coupling, open, repeated washing with water cannon. All the cleaned instruments are dried and soaked in the 1:270 fast enzyme lotion, 2~5min, thoroughly cleaned by hand. Multi enzyme washing pH neutral, mild nature, no corrosion to the endoscope and other instruments. It can effectively decompose the residue on the apparatus of fat, blood, mucus and other pollutants, such as decontamination, descaling, cleaning and other functions. The vibration 40min in the human ultrasonic cleaning machine is cleaned thoroughly by flowing water, pressure water gun and brush. Then dry with dry gauze or dry with pressure air gun. At last, all instruments are soaked in the lubricant of 1:15 for 30s.